The query returns the pulse envelope threshold voltage in the form -6.00000000E+000.
Examples
The following example makes a positive ("on time") width measurement of a pulsed input signal on chan-
nel 3 at a -12 dB threshold .
CONF:PWID:BURS
INP3:BURS:LEV -12
READ?
Typical Response: +1.21513000000000E-006
See Also
INPut3:BURSt:LEVel
226 Keysight 53220A/53230A Programmer's Reference