14 Not Used 16384 (Reserved for future use)
15 Not Used 32768 "0" is returned.
l Once a bit is set, it remains set until cleared by reading the event register or the *CLS (clear status) com-
mand.
l The *RST, SYSTem:PRESet, and STATus:PRESet commands have no effect on this register.
Return Format
The query reads the event register and returns a decimal value which corresponds to the binary-weighted
sum of all bits set in the register. For example, if bit 4 (decimal value = 16) and bit 9 (decimal value = 512)
are set, this command will return +528.
Example
The following example reads the event register (bit 9 is set).
STAT:OPER?
Typical Response: +512
See Also
STATus:OPERation:ENABle
STATus:OPERation:EVENt?
388 Keysight 53220A/53230A Programmer's Reference