EasyManuals Logo

Keysight E8257D Option H1K User Manual

Keysight E8257D Option H1K
20 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #8 background imageLoading...
Page #8 background image
E8257D H1K
Test Procedure
8. Allow an adequate settling time for the DUT and the power sensor before reading. Settling
time for the power sensor is specified for different power levels. The low limit specified for
the power sensor is -45 dBm. Wait 15 seconds before taking a power meter reading when
the DUT power is changed to -45 dBm. After taking the first reading wait 1 second before
taking successive readings. It may be necessary to make repetitive measurements until the
difference between two successive readings is within a specified limit of < 0.01 dB. This limit
must be set so that the range of measurement results is small in comparison to the limit. For
other power levels the wait time before taking a reading is 2 seconds, and the wait time
between successive readings is 0.1 seconds.
9. Compare the Power Level Accuracy to the specification as follows:
Equation 1-1
10. Set the power to +5 dBm and repeat step 8 and step 9.
5
Keysight E8257D Option H1K User's Guide
PLA (P U(P ))-P
dB
gz0
±
gz0
D
UT

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight E8257D Option H1K and is the answer not in the manual?

Keysight E8257D Option H1K Specifications

General IconGeneral
BrandKeysight
ModelE8257D Option H1K
CategoryPortable Generator
LanguageEnglish

Related product manuals