EasyManua.ls Logo

Keysight M8100 Series - Step 15 - Enabling;Disabling the Interleaving Mode in M8199 A Modules

Keysight M8100 Series
72 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
64 M8100 Series Arbitrary Waveform Generators Getting Started Guide
2 Basic Setup for M8100A
Step 15 - Enabling/Disabling the Interleaving Mode in M8199A modules
There are two different interleaving (ILV) modes:
Non interleaved: This mode is active when no combiner serial no. is
registered, and no remote head is connected.
Interleaved: This mode is active when a combiner serial no. is
registered, or a remote head is connected. Interleaved mode is active
when a remote head is connected even though no combiner serial no.
is registered. The connection of a remote head is only checked at
startup.
To enable the ILV, do the following:
1Go to Parameters > Interleaving > Combiner S/N.
2 Enter the serial no. of the ILV option (Combiner) into the Combiner S/N
text field.
3 Restart the M8070B application. After the restart, the ILV mode will be
enabled.
To disable the ILV, do the following:
1Go to Parameters > Interleaving > Combiner S/N.
2 Delete the serial no. in the Combiner S/N text field and press Enter.
3 Restart the M8070B application. After the restart, the ILV mode will be
disabled.
NOTE
The registered combiner serial no. is persistently stored on the module.

Table of Contents