34 Keysight X-Series Signal Generators Error Messages
Error Messages
List of Error Messages
The instrument was configured to use DHCP with a fallback to
Auto-IP to obtain the LAN interface configuration. The instrument
was not able to negotiate with surrounding DHCP servers. Either
DHCP servers are not present, the lease times are less than 5
minutes or they are not allowing this device to obtain an IP
address. Possibly, the LAN cable is disconnected or is broken.
-259 LAN error; The instrument could not obtain a LAN configuration using
DHCP: {name} configuration was used instead. This error occurs if no
DHCP servers are present or the DHCP server did not provide a lease of
at least 5 minutes.
The instrument was configured to use DHCP to obtain the LAN
interface configuration. The instrument was not able to negotiate
with surrounding DHCP servers. Either DHCP servers are not present,
the lease times are less than 5 minutes or they are not allowing
this device to obtain an IP address. Possibly, the LAN cable is
disconnected or is broken.
-259 LAN error; The instrument detected an IP address conflict.
Enter a new IP address or switch to DHCP/AUTO-IP.
-259 LAN error; IP address and default gateway must be on the same subnet.
The instrument automatically changes the offending entries.
The IP address and default gateway are not on the same network as
specified by the subnet mask. Verify that both addresses are on the
same subnet and try again. Contact your network system
administrator for assistance.
-259 LAN error; IP address is inconsistent with subnet mask or IP class