88 Keysight X-Series Signal Generators Error Messages
Error Messages
List of Error Messages
-221 Settings conflict; Real-Time {name} mode was forced off because there
are no longer any formats turned on.
All formats were turned off, so the Real-Time data transform mode
was turned off as well.
-221 Settings conflict; Reference marker not set for marker delta
measurements
No reference marker was set in the analog sweep marker table
editor. A reference marker must be set before marker delta
measurements can be performed.
-221 Settings conflict; Reference marker not set for marker delta
measurements.
Reference marker not set for marker delta measurements.
-221 Settings conflict; The requested operation is not allowed in a two box
{name} setup.
Certain operations, such as setting to equal powers, are not
allowed in a two-ESG setup. A two-ESG setup exists when either
Voltage Other ESG or Percent Power Other ESG are greater than 0.0.
-221 Settings conflict; Resynchronizing to an external data clock can only
be executed when Real-Time Custom mode is on and the BBG data clock is
set to External.
This feature attempts to re-synchronize the system to an external
data clock. This is necessary if the instrument has lost the clock