Keysight X-Series Signal Generators Error Messages 113
Error Messages
List of Error Messages
Frequency (FR) and Amplitude (AP) in compatibility mode. All other
uses of IS are invalid.
-100 Command error; Unable to set new amplitude with UP/DN command.
The instrument is unable to obtain the necessary information to
calculate the new amplitude in compatibility mode with the UP/DN
command.
-100 Command error; Unable to set new frequency with UP/DN command.
The instrument is unable to obtain the necessary information to
calculate the new frequency in compatibility mode with the UP/DN
command.
208 I/O error; Could not access BBG{value}'s PRAM file {name}. The
personality has been turned off.
Deleting AUTOGEN_PRAM files while they are in use can lead to this
error. Turning the personality back on should fix this. If not,
report this error to the nearest sales and service office.
208 I/O error; Unable to read bit file.
There is an error reading the specified bit file. Try selecting the
file again.
208 I/O error; The user file is too large.
The user file selected is too large. Try selecting a smaller file.
DCH files must be less than or equal to 4096 bytes. TPC pattern
files must be less than or equal to 256 bytes.