112 Keysight X-Series Signal Generators Error Messages
Error Messages
List of Error Messages
information in the input string is not valid.
-100 Command error; LO: No low bandwidth ALC mode allowed.
In 8656/57 compatibility mode, LO was received via GPIB. This
command is not supported by the compatibility mode.
-100 Command error; OA is not supported for the current active function.
In 8340/8757 compatibility mode, the OA query was received but it
is not supported for the current active function.
-100 Command error; Query does not contain a valid value.
The data returned by the query does not represent a valid internal
value or hardware setting. The query is provided for backward
compatibility only. The command associated with this query has been
implemented but does not affect the instrument settings .
-100 Command error; R0: No standby mode allowed.
In 8656/57 compatibility mode, R0 was received via GPIB. This
command is not supported by the compatibility mode.
-100 Command error; Remote active function DN/UP not available.
In 8656/57 compatibility mode, either DN or UP was received via
GPIB. These commands are not supported by the compatibility mode.
-100 Command error; Remote active function IS not available.
The instrument only supports the Increment Set (IS) command for