116 Keysight X-Series Signal Generators Error Messages
Error Messages
List of Error Messages
Excessive hardware usage has been detected. Repeated use of the
instrument hardware in the detected manner can cause damage to the
instrument.
226 Media inaccessible; Cannot read or write file {name}. The media is
currently not accessible because a feature that uses it is on (for
example, Streaming or Arb). To fix this, turn off the feature.
A file cannot be opened, read from or written to if a feature that
uses the underlying media is on (for example, Streaming or Arb).
Turning off the feature will fix this.
227 Resampling error; IQ marker resampling failed.
There was a problem resampling the I/Q markers file.
227 Resampling error; IQ resampling ratio out of range: {name}.
The I/Q resampling ratio was out of range. It was either negative,
zero, or too large.
227 Resampling error; IQ waveform resampling failed.
There was a problem resampling the I/Q samples in the waveform
file.
501 Attenuator hold setting over range; Frequency change forced attenuator
adjust.
The firmware has changed the attenuator setting because, while in
attenuator hold mode, a change in frequency setting has forced the
ALC beyond its range.