Keysight X-Series Signal Generators Error Messages 117
Error Messages
List of Error Messages
501 Attenuator hold setting over range; Power set to lower limit.
The firmware has changed the power setting to a value other than
the requested value due to the fact that, while in attenuator hold
mode, the user has requested a power setting that is below the ALC
range for the attenuator setting. The power has been set to the
lower limit.
501 Attenuator hold setting over range; Power set to upper limit.
The firmware has changed the power setting to a value other than
the requested value due to the fact that, while in attenuator hold
mode, the user has requested a power setting that is above the ALC
range for the attenuator setting.
508 Synthesizer Unlocked
The synthesizer is unlocked. Service may be needed.
508 Synthesizer Unlocked; FracN Loop.
The synthesizer is unlocked. Service may be needed.
508 Synthesizer Unlocked; Offset Loop (over).
The synthesizer is unlocked. Service may be needed.
508 Synthesizer Unlocked; Offset Loop (under).
The synthesizer is unlocked. Service may be needed.