517 Calibration failure; A required calibration array was found
uninitialized. {name}
The specified calibration file was found empty or improperly
initialized. Report this error to the nearest sales and service
office
517 Calibration failure; The adjustment DAC does not have the necessary
range to remove the measured error.
517 Calibration failure; Cal aborted by user.
While executing the internal I/Q calibration, either a DCAS was
sent over the GPIB, or the Abort key on the front panel was
pressed.
517 Calibration failure; Calibration Array {name} interpolation failure. A
change in frequency from one calibration point to the next is causing
an interpolation failure because the frequency change is greater than
32.767 MHz.
The cal data interpolation algorithm is optimized to speed
calculations and requires calibration frequency point intervals of
32.767 MHz or less. Intervals greater than 32.767 MHz generate an
interpolation error. Report this error to service office.
517 Calibration failure; Calibration data used as part of an internal
calculation yielded incorrect results. The {name} calibration data is
invalid.
Calibration data used in an internal calculation yielded invalid