List of Error Messages
calibration terms. The calibration data is incorrect and must be
regenerated. Report this error to service office.
517 Calibration failure; The calibration was unable to optimize the
adjustment within a pre-defined number of iterations. Report this
error to service office.
517 Calibration failure; Cannot interpolate the value %1.9e with index
array {name}. Index array vals must be sorted in increasing order.
The specified calibration files have not been properly initialized.
Report this error to the nearest sales and service office
517 Calibration failure; Correction of the measured error requires a DAC
correction greater than the maximum value of the DAC.
While executing an internal calibration, a failure occurred. The
minimization algorithm cannot provide an accurate calibration.
Correction of the measured error requires a DAC correction greater
than the maximum value of the DAC.
517 Calibration failure; DCFM DC overrange.
The instrument was unable to perform a DCFM or DCFM calibration due
to the input signal being outside of the offset range that can be
calibrated for.
517 Calibration failure; The index and data array sizes do not match.
{name} {name}
The specified calibration files have not been properly initialized.
Report this error to the nearest sales and service office