Keysight X-Series Signal Generators Error Messages 121
Error Messages
List of Error Messages
517 Calibration failure; The instrument setting cannot be used in a
recursive manner.
The instrument setting or command cannot be used recursively.
517 Calibration failure; Internal Channel Correction Calibration failure:
{name}
A failure has occurred while executing the Internal Channel
Correction Calibration. The calibration has been aborted.
517 Calibration failure; IQ Calibration failure: {name}
A failure has occurred while executing the internal I/Q
Calibration. The calibration has been aborted.
517 Calibration failure; I/Q Mod Skew Calibration failure: {name}
A failure has occurred while executing the I/Q Mod Skew
Calibration. The calibration has been aborted.
517 Calibration failure; Measurement not stable.
The measurement is not stable and has resulted in an excessive
number of measurement iterations. Report this error to service
office.
517 Calibration failure; The minimization algorithm does not have the
necessary resolution to provide an accurate calibration value. Report
this error to service office.