122                                    Keysight X-Series Signal Generators Error Messages
Error Messages
List of Error Messages
517 Calibration failure; Minimum DAC value exceeded.
            Correction of the measured error requires a DAC correction less
            than the minimum value of the DAC.
517 Calibration failure; Power Verify aborted by user.
            While executing the Power Verify routine, either a DCAS was sent
            over the GPIB, or the Abort key on the front panel was pressed.
517 Calibration failure; Power Verify failure: {name}
            A failure has occurred while executing the Power Verify routine.
            The calibration has been aborted.
517 Calibration failure; The required calibration values could not be read
     from the file.
517 Calibration failure; The required calibration values could not be
     written to the file.
517 Calibration failure; The Skew calibration value is outside the expected
     range. The update to the calibration data has failed.
            The supplied skew calibration value is outside the expected
            adjustment limits. The adjustment limits are 0 to 1000pS. The skew
            cal value will not be applied.
517 Calibration failure; The specified waveform RMS value({name}) is
     incorrect. The RMS value must  be between 0 and the square root of 2
     (1.41421356237...).