Keysight X-Series Signal Generators Error Messages 135
Error Messages
List of Error Messages
Reduce the waveform amplitude or reduce the Runtime Scaling. The
Baseband Generator DAC has a range of +/-32767 however not all of
this range is available for use. The over range message occurs when
the inputs to the D/A converters of the I/Q signals have been
overflowed. Some of the range is required for calibration offsets.
Some of the range is required for overshoot in the Baseband
Generator filters. Typically, a scaling of 70% of the full range
will avoid this error. The widest dynamic range is achieved with
the largest signal that does not cause the error message to occur.
The output signal will be invalid when this error occurs.
629 File format invalid; Import finished with problems: {name} .
This is reported when an invalid list file is loaded.
629 File format invalid; Incorrect sequence file type. {name}
The ARB sequence file operation failed because the specified file
was not a sequence file of the correct type. This can happen when
the user tries to list a single waveform sequence file with the
multi-waveform sequence SCPI query or vice-versa. {name} is the
sequence filename that was specified.
629 File format invalid; Invalid single waveform file selected. Each
segment must have at least 60 samples and segments cannot exceed the
waveform size. Segment #{value} of sequence {name}.
This error is reported when an invalid single waveform sequence is
selected to be played in the ARB system. A valid single waveform
sequence must have a minimum of 60 samples per segment and no
segment may have an offset plus length that exceeds the waveform's