146 Keysight X-Series Signal Generators Error Messages
Error Messages
List of Error Messages
700 State Save Recall Error; Could not save last state file. Setting to
Power On Preset.
The instrument could not save the last state file, so the power-on
behavior has been set to Power On Preset.
700 State Save Recall Error; Failed to recall the instrument state. {name}
The instrument could not recall an instrument state. Previous error
messages may make why this error was reported clear.
700 State Save Recall Error; Failed to save the instrument state. {name}
The instrument could not save an instrument state. Previous error
messages may make why this error was reported clear.
700 State Save Recall Error; Recall data from different FW revision. No
instrument state change. {name}
An attempt was made to recall a state that was saved with an
incompatible version of the instrument firmware. This typically
occurs when a state file is copied from an instrument with a newer
version of firmware to an instrument with an older version of
firmware. Newer versions of instrument firmware can read older
state files. TODO: Parameter is filename.
700 State Save Recall Error; Recall non-saved state register. Recall
ignored.
A recall was attempted for a state register that is unused. If
state file exists, delete explicitly using catalog.