Keysight X-Series Signal Generators Error Messages 15
Error Messages
List of Error Messages
-321 Out of memory; Not enough Waveform Memory on BBG{value}
There is not enough memory left on the baseband board to generate
the required baseband file. Remove any file that might have been
previously downloaded and is not required anymore (e.g. ARB files).
Select different parameters that require a smaller file. Upgrade to
a baseband option with larger memory.
-321 Out of memory; Operation on file: {name}
A file system operation has failed due to a lack of memory in the
system. Reduce memory usage and try again.
-321 Out of memory; The pre-coding bit data for this transport channel is
not generated due to the memory overflow or memory allocation error.
Reduce the amount of coding applied to data or decrease the amount
of data.
-321 Out of memory; Pulse Train subsystem unable to operate.
Operations on a Pulse Train could not be completed because of
insufficient memory.
-321 Out of memory; Shape subsystem unable to operate.
There is insufficient working memory available. Report this error
to service office.
-321 Out of memory; SMI is not functional.