22 Keysight X-Series Signal Generators Error Messages
Error Messages
List of Error Messages
This error cannot be generated under normal operating conditions
and should be reported to service office.
-310 System error; Display could not be initialized: {name}
A fatal error has occurred while the LCD controller was being
initialized. The display is not functional. Repeated errors of this
nature may mean that service is required. Report this error to
service office.
-310 System error; Error backing up system files.
Error encountered while writing system files to backup storage.
-310 System error; Error creating directory: {name}
Error creating a directory. The directory is shown in error
message.
-310 System error; Error restoring system files.
Error encountered while restoring system files from backup storage.
-310 System error; Error sending RS232 XON/XOFF. {name}
Caused by either framing, parity, or overrun errors during RS232
communication.
-310 System error; Error updating RS232 configuration. {name}
Caused by either framing, parity, or overrun errors during RS232
communication.