Keysight X-Series Signal Generators Error Messages 29
Error Messages
List of Error Messages
Informational error that a file system volume was reformatted
because it was damaged or formatting was requested by user.
-310 System error; {name}
An compatibility language specific system error occurred. The
string contains information about the error.
-310 System error; The {name} board's DCM unlock.
The FPGA's DCM reported unlock. The clocks may not work correctly.
Cycle the instrument's power. If the error is repeated, service is
required. Report this error to service office.
-310 System error; {name} in {name} line:{value} file: {name} stack:{name}
An internal error occurred at the specified location within the
specified file.
-310 System error; The {name} option cannot be configured. Required
hardware is missing.
The instrument option requires hardware that was not detected in
the instrument. The hardware must be present to activate the
option.
-310 System error; The {name} subsystem was not found. The requested
operation could not be completed.
The subsystem named in the error was not found. Operations within a
subsystem depend on the subsystem for successful execution. If it