Keysight X-Series Signal Generators Error Messages 31
Error Messages
List of Error Messages
be executed because the string or block contents were too
long for the device to handle (see IEEE 488.2, 10.7.6.1).
-274 Macro parameter error
The macro definition improperly used a macro parameter
placeholder (see IEEE 488.2, 10.7.3).
-273 Illegal macro label
The macro label defined in the *DMC command was a legal string
syntax, but could not be accepted by the device (see
IEEE 488.2, 10.7.3 and 10.7.6.2). For example, the label was
too long, the same as a common command header, or contained
invalid header syntax.
-272 Macro execution error
A syntactically legal macro program data sequence could not
be executed due to an error within the macro definition
(see IEEE 488.2, 10.7.6.3).
-271 Macro syntax error
A syntactically legal macro program data sequence, written
in accordance with IEEE 488.2, 10.7.2, could not be executed
due to a syntax error within the macro definition
(see IEEE 488.2, 10.7.6.3).
-270 Macro error
A macro-related execution error occurred. This error message