Keysight X-Series Signal Generators Error Messages 61
Error Messages
List of Error Messages
Calibration of the connected power sensor is not available while a
measurement is currently running.
-221 Settings conflict; Cannot select and play waveform file. Requires
Option 65x.
Option 65x is required in order to select and play a waveform file.
-221 Settings conflict; Cannot turn AWGN on, option 403 is required for any
AWGN functionality
Option 403 (AWGN) is required to enable the AWGN portion of option
400 (WCDMA).
-221 Settings conflict; cannot update Baseband generator with Even Second
Delay.
There is an error with the CDMA2000 or 1xEV-DV hardware. Try
turning the CDMA2000 or 1xEV-DV mode off and on again to reset the
baseband generator card. If the problem persists, report this error
to service office.
-221 Settings conflict; The carrier BW exceeds the Total BW and has been
reduced by the BW ratio.
The carrier BW exceeds the Total BW and has been reduced by the BW
ratio.
-221 Settings conflict; Carrier formats must be the same.
All digital modulation multicarriers must be the same format.