3 Spectrum Analyzer Mode
3.3 Channel Power Measurement
Command
:CALCulate:CHPower:LIMit:PSDensity:STATe?
Example
:CALC:CHP:LIM:POW:STAT ON
:CALC:CHP:LIM:POW:STAT?
Preset WLAN Mode: ON
All other Modes: OFF
State Saved Yes
Range
OFF|ON
Power Limit Fail (Remote Query Only)
Queries whether a power test passes or fails. When DVB-T radio standard is
selected in SA Mode, the result of this query has no meaning.
Remote
Command
:CALCulate:CHPower:LIMit:POWer:FAIL?
Example
:CALC:CHP:LIM:POW:FAIL?
Notes Query only
When "Power Limit" on page 663 state is OFF, the returned value is always 0 (pass)
When Power Limit state is ON, the returned value is 0 (pass) if power test passes and 1(fail) if power test
fails
PSD Limit Fail (Remote Query only)
Queries whether PSD test passes or fails. When DVB-T radio standard is selected in
SA Mode, the result of this query has no meaning.
Remote
Command
:CALCulate:CHPower:LIMit:PSD:FAIL?
Example
:CALC:CHP:LIM:PSD:FAIL?
Notes Query only
When "PSD Limit" on page 664 state is OFF, the returned value is always 0 (pass)
When PSD Limit state is ON, the returned value is 0 (pass) if PSD test passes and 1(fail) if PSD test fails
3.3.8.5 Advanced
Contains controls for setting advanced functions of the instrument.
Does not appear in VXT.