2 2G0734 June 7, 2007
Safety and Related Information
CAUTION:
Do not use a cell phone within 2 meters of a laser imager. This
proximity includes any imager behind a wall adjacent to your
location.
CAUTION:
Do not use a microwave oven within 4 meters of a laser
imager. Electromagnetic radiation from a microwave oven is
only an issue if after the oven door is closed and latched, the
seal does not maintain an electromagnetic tight fit between the
oven door and oven main housing. Determining if the seal has
an electromagnetic tight fit requires special detection
equipment.
CAUTION:
Do not use in the presence of flammable anesthetics, oxygen,
or nitrous oxide. This equipment does not have a gas-sealed
electronics enclosure and could ignite any flammable or
explosive gases present in its environment.
CAUTION:
This equipment uses a DICOM network port, and is intended
to connect to other medical devices. It is not intended to be
connected directly outside the building. Only an Authorized
Service Provider of Carestream Health, Inc. products may
perform installation and service maintenance.
CAUTION:
This device is NOT to be located within the patient
environment. Therefore, the equipment must not be located
closer than 1.83 meters from a patient bed or chair.
1.83 m