Calibration Procedure
Signature
The contact image sensor (CIS) of the unit is calibrated properly from the factory. With regular, basic use, the
sensor’s sensitivity decreases which can lead to IMAGE errors. To resolve this issue, the unit can be
re-calibrated using the white calibration paper that came with the unit.
3. Then turn the machine backside front. Proceed to pull the top-back cover from the machine upwards.
4. Put the provided laminated calibration card inside the top back door opening until it fits securely against the
hopper rollers.
Please follow the steps below to calibrate the sensor:
1. Ensure the machine is turned ON.
2. On the front of the machine, press the “ADD” and “PRINT” button at
the same time. A calibration screen will display with a message
instructing you to “Open the Cover and put in the calibration paper”.