Do you have a question about the Konica Minolta Ultra KDR and is the answer not in the manual?
Type | Digital Radiography System |
---|---|
Detector Type | Flat Panel Detector |
Detector Size | 17 x 17 inches |
Grayscale | 16-bit |
Exposure Latitude | Wide |
Power Requirements | 100-240 VAC, 50/60 Hz |
Compatibility | DICOM, PACS |
Reference to KDR System Operation Manual for specifications, safety, and handling instructions.
In Ultra, select the Device Configuration tab and configure panel settings.
Prerequisites and initial steps for registering the KDR panel, including device identity list.
Detailed steps for panel registration using the KONICA MINOLTA AeroSDK Setting Tool.
Insert the installation CD and use the Aero Installer to configure the panel.
Verify panel status is 'Connected' in Ultra after acquiring a test patient.
Procedures for calibrating the KDR panel, including preparation and calibration steps.
Remove grid, adjust SID, ensure collimator light coverage, and check X-Ray field.
Exit Ultra, navigate to TechService, select Manual Calibration, and run the calibration sequence.
Test exposures using a test patient to verify calibration; refer to Ultra DR User Quick Guide.