– Floors should be made of wood or concrete or should be covered with
ceramic tiles. If the floor is covered with a synthetic material, the relative
humidity must be at least 30%.
– During operation the unit must therefore not be exposed to extremely
strong magnetic fields.
– Magnetic fields at mains frequency must comply with the typical values
found in business or hospital environments.
ATTENTION! Keep the minimum distance!
The air intake of the Kröber O2 is located on the back side of the unit, the
following is therefore of importance:
– There should be a minimum distance of 30cm from walls, curtains and
other large objects (e.g. cupboards), so that an unrestricted entry of air is
assured at the back of the unit
– Place the Kröber O2 oxygen concentrator in a well ventilated place.
– The Kröber O2 oxygen concentrator must not be used directly beside or
even stacked with other equipment.
ATTENTION! Avoid overheating!
The unit is air cooled, in order to avoid overheating it must not be placed near
heaters, etc.
ATTENTION! Air pollution!
Only install the oxygen concentrator where it is not affected by air pollution or
smoke.
ATTENTION! Ensure simple device isolation!
Place the KröberO2 such that a simple isolation by disconnecting the power
cable from the mains electricity can be carried out!
ATTENTION! No maintenance during operation!
No maintenance or service must be carried out during operation of the
KröberO2!