7.2 Choosing the location
Please consider the following when choosing the location:
– The unit should have 30 cm clearance from any walls, curtains and other large objects
(e.g. cupboards), to ensure unrestricted entry of air through the back of the unit.
– The unit is air cooled. It must therefore not be placed near heaters, etc. At such a
location there is a risk of overheating.
– The unit should be operated only in places where it is not impaired by air pollution or
smoke.
– Always position the device on a horizontal surface.
NOTE!
The unit can be safely lifted and moved by the carrier handle on top of the unit.
ATTENTION! Ensure a sufficient air supply!
Ensure that the air inlet of the Kröber O2 oxygen concentrator is located in a
well-vented spot. Ensure sufficient distance to walls etc.!
The Kröber O2 oxygen concentrator must not be operated directly next to or
even stacked with other equipment. Ensure a sufficient distance from the walls,
etc.! Set up the Kröber O2 oxygen concentrator so that the air inlet and the outlet
are in a well ventilated area.
ATTENTION! Ensure simple device isolation!
Place the KröberO2 such that a simple isolation by disconnecting the power
cable from the mains electricity can be carried out!
ATTENTION! Air pollution!
Only install the oxygen concentrator where it is not affected by airborne
contamination, such as particles and dust, or by smoke! In general, the unit
should be installed in a low-pollutant area.
ATTENTION: Humidity!
Do not install the oxygen concentrator in damp rooms!
ATTENTION: Observe ambient conditions!
When using this device above an altitude of 2000 m above sea level or outside a
temperature of +5 to +40°C or above a relative humidity of 93%, an adverse effect
on the volume flow and the percentage of oxygen is to be expected and
consequently an impairment of the quality results of the therapy!