Kröber O2 - Kröber O2 Vers. 4.0
Operation
86 EN
Alarm type
Display
Description
Power outage
Power failure LED
No LC display
Possible cause:
The power supply to the unit is interrupted.
This leads to an immediate malfunction of the
Kröber O2!
Countermeasures
Is the power cord securely plugged into the
wall outlet?
Has a home fuse been tripped? Check the
fuse.
NOTE!
If a functional test of the power failure
alarm is to be carried out, this can be
done as follows:
Disconnect the mains plug from the socket.
Switch on the unit.
Check the alarm function:
Disconnecting the mains plug during
operation
NOTE!
For safety reasons, an attempt is made
to interrupt oxygen release and oxygen
production.
Mains undervoltage
Possible cause:
The power supply to the unit is insufficient.
Countermeasures
Use of a voltage stabilizer in areas with
known voltage dips in the public supply
Check the alarm function:
Can only be carried out by technicians with
an adjustable power supply.