Kröber O2 - Kröber O2 Vers. 4.0
Operation
88 EN
Alarm type
Display
Description
Volume flow low
Possible cause:
The concentrator cannot deliver the required
volume flow.
Countermeasures
Check whether oxygen hoses are kinked.
Check whether an oxygen hose is blocked
(e.g. by a chair leg or a wheelchair wheel).
Check whether the humidifier has calcified.
Otherwise: Contact the service department.
Check the alarm function:
Keep the outlet connection closed
Volume flow high
Possible cause:
The concentrator delivers more than the
required volume flow.
Countermeasures
Otherwise: The service must be notified.
Check the alarm function:
Only possible by a technician.
WARNING! Risk of health damage!
Before resuming operation of the unit make sure that fault and cause of fault
have been rectified professionally.