Getting Started
Installation
Warnings
Exposure to solvents (i.e. toluene) may degrade the sealing properties of materials used in the front
panel switches.
EXPLOSION HAZARD – Substitution of components may compromise ability to use the unit in Class I,
Division 2 areas.
EXPLOSION HAZARD – DO NOT replace fuses unless the main power has been switched off.
EXPLOSION HAZARD – DO NOT disconnect or disassemble equipment unless the main power has
been switched off or the area is known to be non-hazardous.
DO NOT operate this machine with unbalanced tubes and shields.
Ensure that all collars, cushions and shields are installed properly and in the correct orientation, as
shown in figure 2.
Figure 2: A/B/C/D Cushion Orientation
CUSHION
NOT USED
“A” MODELS
“B” MODELS “C” MODELS “D” MODELS
For best results, the centrifuge should be housed in a controlled environment. It is recommended
that the centrifuge be secured to the benchtop, table, stand etc. using the hardware supplied.
Allow for optimum working clearance around the unit. Mount centrifuge loosely with larger holes
in the tabletop, to allow room for slight movement of centrifuge. Shown in Figures 1-3. Ensure
the unit is level by adjusting the shock mounts.
Figure 1: Schematic of Mounting Bolts
Table Top View (with Zoom)
How to use Mounting Equipment