18
FEATURES: (Con’t)
CONTROL PANEL: (Con’t)
SET CAL CO2 TEMP DOOR SCAN
ºC %CO2 HEAT HOLD HI LIMIT SAMPLE
1 2 3 4 5 6 7 8 9 10 11 12
5. TEMP COMMAND KEY: Pressing the SET key and then the TEMP
command key allows the user with the Up and Down arrow keys to
establish the desired set point temperature in the Incubator.
• High and low limit alarm set points are factory-set at ±1ºC. If the
temperature deviates from these limits, an audible alarm (continuous
beep) will sound.
• While pressing any key will silence the alarm for a one-hour period,
the user should immediately investigate the reason for the alarm and
take steps to correct the problem. Note that indicating light continues
to flash. If problem is not corrected, the alarm will resume after one
hour has passed.
• By pressing the CAL (#3) and TEMP (#5) keys sequentially users can
calibrate temperature. The Up and Down arrow keys will change the
readout to correspond with the readings obtained from known
standards of temperature measurement.
NOTE: DECREASES IN TEMPERATURE AND CO2 DUE TO NORMAL DOOR OPENINGS ARE
TOLERATED WITHOUT INITIATING THE ALARMS.