Connecting to the
Exhaust System
(Optional)
The exhaust system should be installed by a
qualified HVAC contractor.
Exhaust transition kits aid in the removal of
chemicals or applications where a higher degree of
biohazard and particulate removal is required. See
Chapter 7 for ordering exhaust transition kits for the
Purifier HEPA Filtered Enclosure or Purifier Class I
Enclosure. Review Chapter 2 for exhaust
prerequisites and review Chapter 7 for ordering
blower exhaust equipment. See Figure 3-2 for
exhaust kit options on Purifier filtered enclosures.
Consult Labconco Customer Service should you
require help sizing your blower for the exhaust
volume and system static pressure loss.
To ensure compatibility, the selected exhaust duct
material should match the enclosure, procedures
and chemical applications.
!
WARNING: The weight of the exhaust
ductwork system must be supported
independently of the enclosure superstructure
or damage may occur.
!