This Appendix contains technical information about all
the Purifier HEPA Filtered Enclosures including
electrical specifications, environmental operating
conditions and wiring diagrams.
3 Amps, 115V or 2 Amps, 230V, 50/60 Hz, XPert
Filtered Balance System and XPert Filtered
Balance Station.
Environmental Conditions
Indoor use only.
Maximum altitude: 6562 feet (2000 meters).
Ambient temperature range: 41° to 104°F (5° to
40°C).
Maximum relative humidity: 80% for
temperatures up to 88°F (31°C), decreasing
linearly to 50% relative humidity at 104°F
(40°C).
Main supply voltage fluctuations not to exceed
±10% of the nominal voltage.
Transient over-voltages according to Installation
Categories II (Over-voltage Categories per IEC
1010). Temporary voltage spikes on the AC
input line that may be as high as 1500V for
115V models and 2500V for 230V models are
allowed.