Product Service 1-800-522-7658 49
C
C
H
H
A
A
P
P
T
T
E
E
R
R
7
7
T
T
R
R
O
O
U
U
B
B
L
L
E
E
S
S
H
H
O
O
O
O
T
T
I
I
N
N
G
G
Refer to the following table if your Glassware Washer fails to operate properly. If
the suggested corrective actions do not solve your problem, contact Labconco
Product Service for additional assistance.
PROBLEM CAUSE CORRECTIVE ACTION
Detergent cup in
door fails to open
Malfunction of cup Replace cup.
Malfunction of relay Check electrical continuity or replace relay PCB.
Glassware interfering
with cup door
Move glassware.
Liquid detergent
dispense > 0 ml
None
Noises during
operation
Glassware rattling Load glassware properly. Do not allow glassware
to touch.
Door panel vibrating Tighten panel screws.
Wash arms hitting Check glassware loading.
Water inlet line
makes occasional
hammering sound
Check with plumber.
Purified water pump
is activated
No corrective action needed. This is normal for
about 5 minutes while pure water enters the
washer.
Door drops too far
(below door stop)
Bent door hinge Replace door hinge.