39
Laerdal Medical
APPENDICES
M5070A battery
Infant/Child Key 989803139311
category specifications
battery type 9 VDC, 4.2 Ah, lithium manganese dioxide. Disposable, long-life primary cell.
capacity When new, a minimum of 200 shocks or 4 hours of operating time at 25° C.
(IEC 60601-2-4, 2002)
shelf life
(prior to insertion)
A minimum of 5 years from date of manufacture when stored and maintained
according to directions provided in this document.
standby life
(after insertion)
Typically, 4 years when stored and maintained according to directions provided
in this document.
training life Supports 10 hours of use in training mode.
battery limitations Never charge, short circuit, puncture, deform, incinerate, heat above 60° C, or
expose contents to water. Remove the battery when discharged.
environmental qualification
per RTCA/DO-227,
Section 2.3
Meets following acceptance criteria: No leaking, venting, distortion, fire, or
rupture. Change in open circuit voltage <2%.
category specifications
size 16 cm x 6 cm x 0.5 cm.
weight 29 g.
material Polycarbonate.