EasyManuals Logo

Lauda RE 104 User Manual

Default Icon
45 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #26 background imageLoading...
Page #26 background image
Startin
g
u
p
20
6.5.4 Other error messages
Other error messages shown on display.
EEE
changes with
error code, e.g. 0
!
various temperature probe faults.
!
pump fault, proceed as in Section 6.5.3.
If the fault report is repeated
!
pull out the mains plug and try whether the motor
can be rotated by the fan blade inserting a screwdriver into the ventilation opening
at the back of the unit.
Error code 0 ...255
!
microprocessor error.
Error code 1000...1255
!
slave processor error.
Indication can be used for remote diagnosis.
After rectifying the fault, reset with the key.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Lauda RE 104 and is the answer not in the manual?

Lauda RE 104 Specifications

General IconGeneral
Pump capacity flow rate15 L/min
Bath volume4.5 L
Temperature stability±0.01 °C
Circulation pump flow rate15 L/min
Power supply230V 50Hz

Related product manuals