EasyManua.ls Logo

LI-COR LI-6800 - Installation Guide Manual

LI-COR LI-6800
2 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Loading...
LI-6800 6×6 cm Leaf Chamber
The 6 cm Leaf Chamber is for large leaves. This doc-
ument provides installaion instructions for the large leaf
chamber. If the leaf fills less than half the chamber, we
recommend using the 3×3 cm chamber instead.
Detach the chamber from the head
Six screws will be removed to detach a chamber two on
the connector cover, two on the latch mechanism, and two
that hold the chamber to the head.
Caution: Power off the instrument before following these
procedures.
1
Remove the two cover screws and the connector
cover from the bottom of the head.
2
Unplug the cables from the cable bulkhead (i.e.,
LI-190R, leaf temperature thermocouple, and light
source).
3
Position the chamber latch in the parked position and
remove the two latch screws.
4
Loosen both of the captive chamber screws until the
chamber is free of the head.
5
Separate the chamber from the head.
Warning: Do not attempt to operate the mixing fan
without a chamber in place.
While the chamber is removed from the head, inspect the
Air Filters and Dust Trap. Clear any dust or debris before attach-
ing a chamber.
Installation Guide
1

Other manuals for LI-COR LI-6800

Related product manuals