What to do if LiCONiC StoreX STX40 Accessories instrument does not power up?
- DDebra HallAug 9, 2025
If your LiCONiC Accessories instrument doesn't power up, the green light on the power switch is off, and the handler isn't responding to external commands or reacting at all, start by ensuring the main switch at the rear of the instrument is turned on. Then, check the fuse and replace it if damaged (2AT). After that, inspect the voltage setting and the ON/OFF switch. If these steps don't resolve the issue, it could be due to an internal error, and you should call service personnel.