WHAT
DISPLAY SYMBOLS
MEAN
Display
Condition/ Recommended
Symbol
Cause Action
~
Appears during measure- Remain still
and
refrain
ment; blinks
and
beeps from talking; measure-
MEASUREMENT IN
when
pulse is detected.
ment
in
progress.
PROGRESS
00
Battery voltage is
Replace all batteries
too low for monitor with
new
ones.
LOW BATTERY
to
work
properly.
r
Appears
when
systolic
Check
for
air
leakage;
make
,-
,- ,-
and
diastolic measure- sure cuff tube is properly
-
ments are within 10 connected to
monitor
1
mmHg
of each other.
and
exhaust rate is
in
ERROR 1
increments of between
2 mmHg.
and
5
mmHg
r
Appears if pressure Try measurement again
,-
,-
,-
- 2
value is unstable
due
and
remain very still.
to
movement
during
ERROR 2
reading.
r
Appears
when
pressure
Make sure cuff tube is
,-rr
value
did
not
increase properly connected to
-
3
during
cuff inflation:
monitor, cuff is securely
ERROR 3
1)
cuffis
not
fastened fastened
and
exhaust rate
2) air connector is back- is between 2
mmHg
and
wards or
not
connected
5 mmHg. Make sure air
properly.
connector is inserted
NOTE:
If
you
hear
an
announcement
properly into tube
indicating
that
a
measurement
error
and
monitor.
has occurred,
take
the
corrective
MONITOR
-
~
-
CUFF
HOSE
actions
shown
above
and
re-measure.
3
0
LifeSource
Health
Line (Toll-Free):
1-888-726-9966

Other manuals for LifeSource UA-767T

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the LifeSource UA-767T and is the answer not in the manual?

LifeSource UA-767T Specifications

General IconGeneral
BrandLifeSource
ModelUA-767T
CategoryBlood Pressure Monitor
LanguageEnglish

Related product manuals