9
Function Range Resolution Accuracy Notification
Temperature
-40
~
1000°C
-40~ 40°C
1°C
±4°C
K thermocouple is only
applicable for temperature less
than 250°C
/
482˚F.
>40~500°C ± (1.0%+4)
>500~1000°C ± (2. 0%+4)
Temperature
-40
~
1832°F
-40~104°F
1°F
±5°F
>104~932°F ± (1.5%+5)
>932~1832°F ± (2.5%+5)
DC Current
2000
μ
A 1
μ
A ± (1.0%+2) When inupt current > 10 A, "OL"
symbol appears and buzzer
beeps.
μ
A mA range:
F1 Fuse 0.2A
/
250V
Ф
5
×
20mm.
10 A range: F2 Fuse 10 A/250 V
Ф
5
×
20 mm.
20.00 mA 0.01 mA ± (1.0%+2)
200.0 mA 0.1 mA ± (1.0%+2)
10.00 A 0.01 A ± (1.2%+5)
Continuity
diode
0.1
Ω
If the measured resistance is
greater than 50
Ω
, the measured
circuit will be regarded as in
open status, and the buzzer
turns off.
If the measured resistance is
less than 10
Ω
, the measured
circuit will be regarded as in
good conduction status, and the
buzzer beeps.
0.001 V
Open circuit voltage: 2.1 V, test
current is about 1 mA.
Silicon PN junction voltage is
about 0.5-0.8 V.
Resistance
200.0
Ω
0.1
Ω
± (1.0%+2)
Measurement result = reading of
resistor - reading of shorted test
leads.
2000
Ω
1
Ω
± (0.8%+2)
20.00 k
Ω
0.01 k
Ω
± (0.8%+2)
200.0 k
Ω
0.1 k
Ω
± (0.8%+2)
20.00 M
Ω
0.01 M
Ω
± (1.2%+3)
GB