EasyManua.ls Logo

LIMIT 7000 - Page 7

LIMIT 7000
100 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
English
5
30 groups of measurings can be stored.
1. When M is showed on the display push SAVE button.
The Display show number of stored values.
2. Push READ button for reading stored values. R
shows on the display and latest value are showed. Push
▲▼- buttons for reading the other stored values.
3. Push DEL/MENU-button to delete stored value on the
display.
4. Return by push FUNCTION-button.
Calibration
Standard method for calibration requires an external
calibrator. Code no 15152-0301.
1. Push the Ө-button.
2. Select A-weighing and SLOW.
3. Place the microphone into the calibrator. Turn the
calibrator on.
4. Adjust the meters CAL potentiometer on the meters
right side with a small screwdriver so the value on the
display matches the output of the calibrator 94 dB or
114 dB.
Limit 7000 have a built in signal for calibration.
1. Push FUNCTION-button till CAL shows on the display.
2. The value on the display should be 94. If not, adjust
the meters CAL potentiometer on the meters right side
with a small screwdriver so the value on the display is
94 dB.

Related product manuals