This procedure must be carried out when both the AlcoCal
®
and the
SD-400 are at room temperature in the range 15 – 30°C:
1. Prepare the AlcoCal
®
standard
Note the value of the AlcoCal
®
to be used (making
allowance for atmospheric pressure and units of
measurement, as previously described). Now fit the
Dispensing Top to the AlcoCal
®
-M, or the Connector Tube
to the regulator of the AlcoCal
®
-2AL. If the AlcoCal
®
-2AL
has not been used within the previous two hours, press and
hold down the push-button for about five seconds to flush
out of the regulator of any stale and depleted alcohol vapour.
2. Prepare the SD-400 Instrument
First, ensure this has not been used at all in the last hour.
Now put it into its Calibration Adjustment Mode. This
ensures the SD-400 is in its correct temperature range;
allows an actual resetting of the instrument’s analytical
function to be made; and ensures the event is not stored in
the subject test memory. So insert the Dongle into the
Communications Port on the side of the instrument.
Now switch the instrument on and then IMMEDIATELY
press and hold down C. This MUST be done BEFORE the
diagnostics sequence is complete. When a continuous tone
is heard, release C and wait for the READY light to flash [the
display now shows the software version (eg 3.9)]. When the
READY light flashes, press and release D. The display now