2.2 Specifications
Operating temp
0˚C to +50˚C (32˚F to 120
With water cooling: 0˚C to 120˚C (32˚F to 250˚F)
LED Indicators
Green = power on, Red = target detect, Orange=
Detector signal
saturation or dirty lens, Off=Alarm
Output
closing: 7.5ms, opening: 3ms
DC 24V PNP,NPN Transistor output: max 100mA
Diagnostic outputs: control, alarm, signal amplitude
24 VDC, 110VAC, 220VAC @ 30VA
conductor TEFLON cable with
Insulation 20MΩ, 500VDC, ≥500˚C (930˚F)
Housing
Hermetically sealed cast aluminum enclosure rated IP66, includes cooling
water connections and air purge.
mm H (with mounting bracket)
(12.4”L x 4.3”W x 10.4”H)
Water Cooling
25
˚
C, 4 bars pressure (60 psi), 1-2L/min (0.26 to 0.52 USG),
¼” flexible tube
Air Purging
50 to 200 g/cm
(0.7 to 2.8 psi), 4-16L/min (0.14 to 0.56 cfm)
¼” compression fitting
Table 2
2.3 Operating Principle
Detector- The HMD-SM uses a Lead Sulfide (PbS) photocell, which is characterized by a very
strong signal output for the range of infrared energy emitted by steel, compared with other
detector materials. The high signal-to-noise ratio from this detector virtually eliminates
detection errors. Figure 2.1 shows the various types of photocell spectrums and the
infrared emission spectrum of steel in 250
˚
C and 1000
˚
C.