LRF-3000S Ultrasonic Transit-time Flow meter
- 48 -
RT-(cr)(lf)
Return negative accumulative flow
±ddddddd.d±d(cr)
RTN(cr)(lf)
Return net accumulative flow
±ddddddd.d±d(cr)
RTH(cr)(lf)
Return net accumulative energy(hot)
±ddddddd.d±d(cr)
RT-(cr)(lf)
Return net accumulative energy(cold)
±ddddddd.d±d(cr)
RER(cr)(lf)
Return instantaneous energy value
±d.ddddddE±dd(cr)
RA1(cr)(lf)
Return analog input value of AI1
(Temperature, Pressure, etc.)
±d.ddddddE±dd(cr)
RA2(cr)(lf)
Return analog input value of AI2
(Temperature, Pressure, etc.)
±d.ddddddE±dd(cr)
RID(cr)(lf)
Return Net address of the instrument
ddddd(cr) 5 bits in length
RSS(cr)(lf)
Return signal intensity
UP:dd.dDN:dd.dQ=dd(cr)
REC(cr)(lf)
Return current error code
*R/*D/*E Note 3
RRS(cr)(lf)
Return Relay Status
ON/OFF(cr)
RDT(cr)(lf)
Current date and time
yy-mm-ddhh:mm:ss(cr)
RSN(cr)(lf)
Return serial number
dddddddt(cr) Note 4
SFQdddd.d(cr)(lf)
OCT setting
dddd.d(cr)
Successful setting will back to “OK”
SCLdd.d(cr)(lf)
Current setting
dd.d(cr)
Successful setting will back to “OK”
SRS(cr)(lf)
Start quantitative control
OK(cr)
Successful setting will back to “OK”
P
Prefix of return command with check
Note 5
W
Networking command prefix of numeric
string address
Note 6
Note:
1. (cr)expresses carriage return. Its ASCII value is 0DH. (lf) expresses line feed. Its ASCII value is 0AH.
2. "d" expresses 0-9 number. 0 value is expressed as +0.000000E+00.
3. "d" expresses 0-9 numbers. There is no decimal point in integral part before "E".
4. dddddddd means the serial number of the instrument, t means the model of the instrument.
5. The character P can be added before every basic command. It means that the transferred data has CRC