26
equipment does cause harmful interference to radio or television reception, which
can be determined by turning the equipment off and on, the user is encouraged to
try to correct the interference by one or more of the following measures:
• Reorient or relocate the receiving antenna.
Increase the separaon between the equipment and receiver.
Connect the equipment into an outlet on a circuit dierent from that to which
the receiver is connected.
• Consult the dealer or an experienced radio/TV technician for help.
Note:
The Upper Arm Blood Pressure Monitor LK-A2 meets the electromagnetic
compability requirements of YY0505, YY0670, IEC60601-1-2;
Users should install and use the electromagnetic compatibility information
provided by the random files. Portable and mobile RF communication devices
may affect the performance of the Upper Arm Blood Pressure Monitor LK-A2.
Avoiding strong electromagnec interference when used, such as Close to mobile
phones, microwave ovens, etc.
The instructions for the guide and the manufacturer are detailed in the
aachment.
Warning:
Use of this equipment adjacent to or stacked with other equipment should be
avoided because it could result in improper operaon. If such use is necessary,
this equipment and the other equipment should be observed to verify that they
are operang normally.
Use of accessories, transducers and cables other than those specied or provided
by the manufacturer of this the LK-A2 could result in increased electromagnec
emissions or decreased electromagnec immunity of this equipment and result
in improper operaon.
Portable RF communicaons equipment (including peripherals such as antenna
cables and external antennas) should be used no closer than 30 cm (12 inches) to
any part of the LK-A2, including cables specied by the manufacturer. Otherwise,
degradaon of the performance of this equipment could result.
Aachment:
Guidance and manufacturer´s declaraon – electromagnec emission – for all