27
EQUIPMENT AND SYSTEMS
Guidance and manufacturer´s declaraon – electromagnec emission
The LK-A2 is intended for use in the electromagnec environment specied below. The customer
or the user of LK-A2 should assure that it is used in such an environment.
Emissions test Compliance Electromagnec environment-guidance
RF emissions
CISPR 11
Group 1
The LK-A2 uses RF energy only for its internal
funcon. There for, its RF emissions are very
low and are not likely to cause any interfer-
ence in nearby electronic equipment.
RF emissions
CISPR 11
Class B
The LK-A2 suitable for use in all establish-
ments, including domesc establishments and
those directly
connected to the public low-voltage power
supply network that supplies buildings used
for domesc purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage uctuaons
icker emissions
IEC 61000-3-3
Complies
Guidance and manufacturer's declaraon – electromagnec immunity – for all
EQUIPMENT and SYSTEMS
Guidance and manufacturer´s declaraon – electromagnec immunity
The LK-A2 is intended for use in the electromagnec environment specied below. The customer
or the user of the LK-A2 should assure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance level
Electromagnec environment
-guidance
Electrostac
discharge (ESD)
IEC 61000-4-2
± 8 kV contact
± 2 kV, ± 4 kV,
± 8 kV,
± 15 kV air
± 8 kV contact
± 2 kV, ± 4 kV, ± 8
kV,
± 15 kV air
Floors should be wood,
concrete or
ceramic le. If oors are
covered with
synthec material, the rela-
ve humidity
should be at least 30 %.