A
Anna BarreraSep 9, 2025
What does ERROR 2 mean on my Lowenstein Medical Diagnostic Equipment?
- FfergusonbarbaraSep 9, 2025
If you're seeing ERROR 2, it indicates an issue with the RAM. Return the device to the manufacturer to have it checked.
What does ERROR 2 mean on my Lowenstein Medical Diagnostic Equipment?
If you're seeing ERROR 2, it indicates an issue with the RAM. Return the device to the manufacturer to have it checked.
How to fix a Lowenstein Medical SOMNOcheck micro Diagnostic Equipment that does not switch on?
If your device isn't turning on, try these steps: * Replace the batteries or charge the rechargeable batteries, as they may be discharged. * Ensure the batteries or rechargeable batteries are inserted with the correct polarity. If neither of these steps works, the internal fuse may have tripped, and the device should be returned to the manufacturer for repair.
Why are the batteries/rechargeable batteries supplying no power to my Lowenstein Medical Diagnostic Equipment?
If the batteries or rechargeable batteries aren't supplying power, here's what might be happening: * The contact springs may not be making proper contact. Try reinserting the batteries or rechargeable batteries. * The batteries or rechargeable batteries themselves may be defective. Try replacing them. * The batteries or rechargeable batteries may be discharged. Replace the batteries or charge rechargeable batteries.
What to do if my Lowenstein Medical Diagnostic Equipment is displaying implausible values?
If the results are displaying implausible values, it could be due to a couple of reasons: * The plug connections may not be properly connected. Ensure the connections are firm and check the sensors. * The sensor may have gotten dirty or slipped out of position during the measurement. In this case, repeat the recording with clean or new sensors.
How do I troubleshoot connection issues with my Lowenstein Medical SOMNOcheck micro Diagnostic Equipment?
If you're having trouble connecting to the device, consider the following: * Check all cable connections to ensure they are secure and not interrupted. * If the device has switched off, turn it back on using the power button. The device may automatically switch off if the connection has been inactive for more than 5 minutes.
What causes a Lowenstein Medical Diagnostic Equipment to display no signals and how can I fix it?
If the device is displaying no signals, it's possible that the plug connections are not properly connected. Ensure that all connections are firm.
What does ERROR 102 mean on my Lowenstein Medical SOMNOcheck micro?
If you're seeing ERROR 102, it indicates an issue with the EEPROM. Return the device to the manufacturer to have it checked.
What does ERROR 1 mean on my Lowenstein Medical SOMNOcheck micro Diagnostic Equipment?
If you're seeing ERROR 1, it indicates a program code issue. Return the device to the manufacturer to have it checked.
What does ERROR 3 mean on my Lowenstein Medical Diagnostic Equipment?
If you're seeing ERROR 3, it indicates an issue with the UART. Return the device to the manufacturer to have it checked.
What does ERROR 5 mean on my Lowenstein Medical Diagnostic Equipment?
If you're seeing ERROR 5, it indicates an issue with the SPI. Return the device to the manufacturer to have it checked.
Data Storage | Internal memory |
---|---|
Connectivity | USB |
Channels | 3 |
Device Type | Sleep Diagnostic Device |
Parameters Measured | Pulse, Oxygen Saturation |
Data Transmission | USB |
Sampling Rate | 1 Hz |
Operating Temperature | 10°C to 40°C |
Storage Temperature | -20°C to 60°C |
Humidity | 10% to 95% relative humidity (non-condensing) |
Important warnings and cautions for safe operation and handling of the device and its parts.
Step-by-step guide to attach the device and sensors to the patient.
Procedures for preparing, starting, and managing data recordings.
Steps for visual inspection and functional tests to ensure device readiness.
Identifies common device faults, error messages, and their remedies.