Technical Specifications
Outputs Sinusoidal AC
0 – 200 Volts peak to peak
0 – 60 mA RMS
Delivered current displayed for each output
Frequency 4000 – 4250Hz continuous or burst modulated from 1 – 250 with a 50% duty cycle.
Vector Sweep provides amplitude modulation of Output 1 and 2 from -20% to preset
intensity over an eight second period.
Frequency Adjust provides selectable output frequency of 4000-4250Hz.
Timer Digital countdown from 1 – 98 minutes with interrupt and reset capability.
Safety Stimulation output is inhibited each time the STOP key is actuated. Output is automatically
reset to 0 each time the patient remote stop is depressed or when the timer counts down to
no time remaining.
Line Leakage Patient cables to line - less than 10 uA.
Accessories 1 - Red 6 DIN to two .080 pin leadwires
1 - Black 6 DIN to two .080 pin leadwires
6 - packages of 2 x 2 LSI Silver Tab Repositionable Electrodes
6 - packages of 2 x 4 LSI Silver Tab Repositionable Electrodes
1 - line cord
2 - remote stop
1 - operator manual
1 - pad placement
1 - quick reference guide
1 - training video