D - PREMOD: When off (no light), channel 1 will operate at an uninterrupted 4000Hz and channel 2 will
operate at 4000Hz plus the frequency selected on function A (Frequency Readout). When on (light on),
both channels will operate at a carrier frequency of 4000Hz and both channels will be synchronized to
deliver the frequency as shown in function A.
E - SWEEP: Frequency will sweep between the range indicated in function B ( Frequency Selection).
Press key pad to activate light and this function.
F - SET: This key is used for a special programming option. Please refer to page 10.
G - TIMER: Indicates treatment time remaining.
H - TIMER SET: Set key automatically enters 10 minutes to timer.
I - TIME ADJUST: By pressing the up or down key, the treatment time may be raised or lowered from the
10 minutes which is set automatically.
J - START: Activates timer and outputs to channels 1 and 2.
K - STOP: Stops timer and output to channels 1 and 2.
L - MILLIAMPS OUTPUT: Indicates the output, measured in milliamps, which is actually being delivered
to the tissue.
M - CURRENT OUTPUT CONTROLS: Keypads will increase or decrease the delivered current to channels
1 and 2.
N - 2500HZ MODE: Automatically activates 50Hz premodulated setting typically used for Russian Stimu-
lation and muscle spasm reduction. When using the 2500Hz setting, SWEEP and VECTOR controls will
be in the off position. For simplication, these features are programmed in a locked off mode.
O - 10/10 MODE: Activates the 10 second on and 10 second off function. When activated, the off portion
of the cycle will precede the on portion. Automatic ramping of 2.5 seconds up and 2.0 seconds down is
provided.
P - 10/50 MODE: Activates the 10 second on and 50 second off function. When activated, the off portion
of the cycle will precede the on portion. Automatic ramping of 2.5 seconds up and 2.0 seconds down is
provided.
Q - VECTOR: When activated, channel 1 and 2 will alternately drop 20% from their preset output and
return to the preset gure.
Notes Concerning Functions M, N, O and P: As a safety feature, output cannot be changed while the
10/10 or 10/50 functions are in the off cycle or when the VECTOR light is on. Also, since there is no reason
to utilize the VECTOR while the 10/10 or 10/50 functions are in use, the unit will not allow dual activation.
R-
REMOTE STOP: Depressing the button on the remote control will stop all output of the unit. A repeating
tone will sound, signaling the operator that the patient has activated the switch.
S - OUTPUT JACKS: Output jacks for channel 1 and 2.
***Complete operating instructions with examples are reviewed in the Instructional Video (CD-ROM) provided with your unit.***

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the LSI System II and is the answer not in the manual?

LSI System II Specifications

General IconGeneral
BrandLSI
ModelSystem II
CategoryTest Equipment
LanguageEnglish