EasyManua.ls Logo

Luwa HTT-V Series - Page 15

Luwa HTT-V Series
18 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operating Instruction
Q5.25.45
TLT Page
15 / 18
Draft created/changed: 31.01.03/TT Document created/changed: 03.08.2012 / Rea
Revision: - Revision: -
Draft released: 18.02.03/DEL Document released: 03.08.2012 / Rea
Jede unerlaubte Verwendung ist strafbar / Any unpermitted use is subject to prosecution.
7
7
.
.
7
7
O
O
f
f
f
f
s
s
e
e
t
t
a
a
d
d
j
j
u
u
s
s
t
t
m
m
e
e
n
n
t
t
f
f
o
o
r
r
a
a
n
n
a
a
l
l
o
o
g
g
u
u
e
e
o
o
u
u
t
t
p
p
u
u
t
t
:
:
With this function, the analogue output and the value for manual output can be adjusted
from -5%RH to 105%RH. This allows an adaption to inaccurate analogue inputs.
7.7.1 Settings before correction
Adjust the desired values for Range L and Range H at the Transmitter HTT-V,
for example 0%RH (RLs) and 100%RH (RHs).
Set the desired output range (0/2..10V, 0/4..20mA).
Set the control system in the same way.
7.7.2 Measurement
Enter a manual value for example 10%RH (S1). Note the range on the control system, e.g. 11%RH (M1).
Enter a manual value e.g. 90%RH (S2). Note the range on the control system e.g. 92%RH (M2)
7.7.3 Calculation Gain and Offset
7.7.4 Calculation corrected settings Range L and Range H
To adjust 0.9[%RH]
To adjust 102.1[%RH]
The display on the control system should now concur with the one on the transmitter. The linearity of both sys-
tems is assumed.
Should the deviation at the nominal value of the measurement of the system be too big due to a nonlinearity,
chose for S
2
a value around the nominal measuring value and for S
1
the RLs value.